מכשיר מעבדתי – מסחרר טרמי
Huber MPC-205 Heating Circulator
(מסחרר טרמי (חימום
MPC-Controller:
Modern and easy to use microprocessor controller with a large temperature display.
Limited to essential functions only:
- Simple operation using only 3 keys
- Large temperature display, actual temperature and set point
- LED indicators for pump, cooling and heating
- RS232/serial with the LAI commands G,v,L; (SpyLight compatible)
Technical Data↓
- Operating temperature range 45…200 °C
- Minimum temperature with water cooling 20 °C
- Minimum temperature with refrigeration -30 °C
- Temperature stability at 70°C 0,05 K
- Temperature adjustment digital
- Temperature indication digital
- Internal temperature sensor Pt100
- Safety classification Class III / FL
- Heating power 2 kW
- Pressure pump yes
- max. delivery 20 l/min
- max. delivery pressure 0.2 bar
- Suction pump yes
- max. delivery (suction) 17 l/min
- max. delivery pressure (suction) 0,18 bar
- Pump connection M16x1 male
- Bath volume 5 l
- Width bath opening WxD 105×90 mm
- Bath depth 150 mm
- Overall dimensions WxDxH ** 178x337x355 mm
- Height of bath opening 190 mm
- Net weight 9 kg
- Power supply requirement 230V 1~ 50/60Hz
- max. current 10 A
- Fuse (1 phase) 10 A
- min. ambient temperature 5 °C
- max. ambient temperature 40 °C
Cooling coil for (tap) water (3/8"). With adjustable overtemperature protection according to DIN 12876
Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).
It is recommended that water is mixed with either glysantin or ethylene glycol for freeze protection.
A more resistant plastic is available on request at an additional cost.
MPC-Controller:
Modern and easy to use microprocessor controller with a large temperature display.
Limited to essential functions only:
- Simple operation using only 3 keys
- Large temperature display, actual temperature and set point
- LED indicators for pump, cooling and heating
- RS232/serial with the LAI commands G,v,L; (SpyLight compatible)
- Accessories and periphery: Adapter nom. dia 12 *, dummy plugs*, drain valve, sleeve nuts thread M16x1*, bath cover*
- standard equipment
- Output data valid for: Room temperature 20°C
In accordance with EN60034-1 the following voltage and frequency tolerances are valid:
Voltage + / – 10%, as long as the frequency tolerance does not run in the opposite direction.
Example: -10% voltage and +3% frequency -> not allowed !
-10% voltage and -3% frequency -> allowed.
** Please respect space requirements. See operating conditions at